Model 6220/6221 Reference Manual
Remote Operations 10-13
Return to
<name>
Name parameter
— Select a parameter name from a listed
group.
<name> = MXB
= RECiprocal
:CALCulate:FORMat MXB
Select Mx + B calculation
<NRf>
Numeric representation format
— A number that can be
expressed as an integer (e.g., 8), a real number (e.g., 23.6), or
an exponent (e.g., 2.3E6).
:SOURce:SWEep:COUNt 10
Set sweep count to 10
<NDN>
Non-decimal numeric
— A non-decimal value that can be
used to program status enable registers. A unique header
identifies the format: #B (binary), #H (hexadecimal), and #Q
(octal).
*SRE #B10001
Set bits B0 and B4 of Service
Request Enable Register
<n>
Numeric value
— Can consist of an NRf number or one of the
following name parameters: DEFault, MINimum, or MAXimum.
When the DEFault parameter is used, the instrument is pro-
grammed to the *RST default value. When the MINimum
parameter is used, the instrument is programmed to the low-
est allowable value. When the MAXimum parameter is used,
the instrument is programmed to the largest allowable value.
:ARM:TIMer 0.1
Sets timer to 100 msec.
:ARM:TIMer DEFault
Sets timer to 0.1 sec.
:ARM:TIMer MINimum
Sets timer to 0s.
:ARM:TIMer MAXimum
Sets timer to 99999.99 sec.
•
Angle brackets
< > — Used to denote a parameter type. Do not include
the brackets in the program message.
:DISPlay:ENABle <b>
The <b> indicates that a Boolean type parameter is required. Thus, to
enable the display, you must send the command with the ON or 1 parame-
ter as follows.
:DISPlay:ENABle ON or 1
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...