8-10
Triggering
Model 6220/6221 Reference Manual
Return to
SCPI commands — triggering
Triggering commands are listed in
. Additional information for each com-
mand is provided in notes that follow the tables.
Table 8-2
Trigger commands
Command*
Description
Default
INITiate
Initiate one trigger cycle.
1
OFF
ABORt
Reset trigger system.
2
Arm layer commands:
ARM:SOURce <name>
Select event detector control source.
3
<name> = IMMediate, TIMer, BUS,
TLINk, NSTest, PSTest, NSTest,
or MANual.
IMM
ARM:SIGNal
Bypass ARM control source.
4
ARM:TIMer <n>
Set timer interval (0 to 99999.99 sec).
1 msec resolution.
5
0.1
ARM:DIRection <name>
Control ARM source bypass.
6
<name> = SOURce or ACCeptor
ACC
ARM:ILINe <NRf>
Set ARM layer input signal line (1-6).
7
1
ARM:OLINe <NRf>
Set ARM layer output signal line (1-6).
8
2
ARM:OUTPut <name>
Enable/disable ARM layer output trigger.
9
<name> = TENTer, TEXit, or NONE
NONE
Trigger layer commands:
TRIGger:SOURce <name>
Select event detector control source.
10
<name> = IMMediate or TLINk.
IMM
TRIGger:SIGNal
Bypass TRIG control layer.
11
TRIGger:DIRection <name>
Control TRIG source bypass.
12
<name> = SOURce or ACCeptor
ACC
TRIGger:ILINe <NRf>
Set TRIG layer input signal line (1-6).
13
1
TRIGger:OLINe <NRf>
Set TRIG layer output signal line (1-6).
14
2
TRIGger:OUTPut <name>
Select TRIG layer output trigger.
15
<name> = SOURce[1], DELay, or NONE
NONE
* Arm layer commands control entire sweep. Trigger layer commands control individual sweep steps.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...