Model 6220/6221 Reference Manual
SCPI Reference Tables
14-5
Return to
Section 14 topics
Table 14-2
Display command summary
Command
Description
Default
Sec
SCPI
DISPlay
Display control commands:
:ENABle <b>
Enables or disables front panel display and controls. (Note)
√
:ENABle?
Queries state (on or off) of the display and controls.
√
[:WINDow[1]]
Top line display characters:
√
:TEXT
Text message:
√
[:DATA] <a>
Define ASCII message “a” (up to 20 characters).
(Note)
√
[:DATA]?
Read text message
√
:STATe <b>
Enable or disable text message.
(Note)
√
:STATe?
Query state (on or off) of text message.
√
:ATTRibutes?
Query attributes of message characters:
Blinking (1) or not blinking (0)
:WINDow2
Bottom line display characters:
√
:TEXT
Text message:
√
[:DATA] <a>
Define ASCII message “a” (up to 32 characters).
(Note)
√
[:DATA]?
Read text message
√
:STATe <b>
Enable or disable text message.
(Note)
√
:STATe?
Query state (on or off) of text message.
√
:ATTRibutes?
Query attributes of message characters:
Blinking (1) or not blinking (0)
Note: *RST and SYSTem:PRESet have no effect on the display circuitry, a text message, or the state of the mes-
sage modes. Pressing
LOCAL
or cycling power enables (ON) the display circuitry, cancels all text mes-
sages and disables (OFF) the message modes. Pressing the
DISP
key will also enable the display.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...