Model 6220/6221 Reference Manual
Averaging Filter, Math, and Buffer
6-15
Return to
Figure 6-3
Buffer recall
For
remote programming
, the commands to recall delta readings and statistics
. The following example demonstrates the command
sequence read buffer readings and statistics.
Example
– Request all stored delta readings, and calculate and return the mean
average:
Remote programming – Buffer
The commands associated with buffer operation are listed in
.
TRACe:DATA?
‘ Request all stored delta readings.
CALC2:FORMat MEAN
‘ Select the mean buffer calculation.
CALC2:STATe ON
‘ Enable buffer calculation.
CALC2:IMMediate
‘ Perform the mean calculation.
CALC2:DATA?
‘ Request the result of the mean calculation.
PUSH TO ENTER
+123.45678mV
00000
Isrc +1.0000mA
Time 000000.001 s
Reading
Reading Number
Current Source
Timestamp
Buffer Readings
RANGE
RANGE
or
RANGE
RANGE
or
A) Front panel edit keys to display buffer readings:
Model 6220:
Model 6221:
EDIT/
LOCAL
B) Use EDIT/LOCAL key to display buffer statistics:
Min
®
Max
®
Pk-Pk
®
Avg
®
Std Dev
®
Buffer Readings