14-14 SCPI Reference Tables
Model 6220/6221 Reference Manual
Return to
Section 14 topics
Table 14-8
System command summary
Command
Description
Default
Sec
SCPI
:SYSTem
:COMMunicate
Communication interfaces:
:SELect <name>
Select interface: SERial, GPIB or ETHernet.
:GPIB
GPIB interface:
:ADDRess <NRf>
Set primary address: 0 to 30.
:ADDRess?
Query primary address.
:SERial
RS-232 interface:
:CONTrol
Handshaking:
:RTS <name>
Select handshaking: IBFull, RFR, or OFF.
:RTS?
Query handshaking.
:PACE <name>
Set flow control: XON or OFF
:PACE?
Query flow control.
:TERMinator <name>
Set output terminator: CR, LF, CRLF, or
LFCR.
:TERMinator?
Query output terminator.
:BAUD <n>
Set baud rate: 300, 600, 1200, 2400, 4800,
9600, 19.2k, 38.4k, 57.6k, or 115.2k.
:BAUD?
Query baud rate.
:SEND <data>
Send data via RS-232.
:ENTer?
Reads data from serial port.
:LOCal
RS-232 only – Take 622x out of remote.
:REMote
RS-232 only – Put 622x in remote.
:RWLock <b>
RS-232 only – Enable or disable local lockout.
:RWLock?
RS-232 only – Query state of local lockout.
:ETHernet
Ethernet interface (6221 only):
:ADDRess <string>
Set IP address: “n.n.n.n”.
:ADDRess?
Query IP address.
:MASK <string>
Set subnet mask: “n.n.n.n”.
:MASK?
Query subnet mask.
:GATeway <string>
Set Ethernet gateway: ”n.n.n.n”.
:GATeway?
Query Ethernet gateway.
:DHCP <b>
Enable or disable DHCP.
:DHCP?
Query state (on or off) of DHCP.
:SAVE
Saves Ethernet setting changes.
:KEY <n>
Simulate key press.
See
“SYSTem key-press codes,” on page 13-9
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...