5-30
Delta, Pulse Delta, and Differential Conductance
Model 6220/6221 Reference Manual
Return to
6.
Read Delta readings
– While Delta is running, the latest Model 2182/
2182A Delta reading can be read by the Model 622x using the following
command:
The above read command reads the last Delta reading that was performed
by the Model 2182/2182A. If this command is sent before a new reading is
available, the last Delta reading will again be returned.
7.
When finished with Delta, it can be disarmed by sending the following command:
8.
Recall stored Delta readings
– Model 2182A Delta readings were sent to
the buffer of the Model 622x. Send the following read command to read the
buffer:
Setup commands
Delta setup and arm commands
Commands to set up and arm Delta are listed in
.
SENS:DATA?
‘ Reads the latest Delta reading.
NOTE
If a read command is sent when Delta is not run-
ning, error -221 Settings Conflict will occur. See
for details on errors associated with
Delta operation.
NOTE
Details on using Model 622x “
”
are provided on
.
SOUR:SWE:ABOR
‘
‘
Stops Delta and places the Model 2182A in the
local mode.
TRACe:DATA?
‘ Read Delta readings stored in 622x buffer.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...