35
:CLEar
[:IMMediate]
:AUTO <b>
:AUTO?
:IMMediate
:LIMit 2
:UPPer
[:DATA] <n>
[:DATA]?
:LOWer
[DATA] <n>
[DATA]?
:STATe <b>
:STATe?
:FAIL?
:CLEAR
[:IMMediate]
:AUTO <b>
:AUTO?
:IMMediate
Path to clear failed test:
Clear failed test indication.
Enable or disable auto clear.
Query auto clear.
Re-perform limit tests.
Path to control LIMIT 2 test:
Path to configure upper limit:
Set upper limit (-100e6 to
100e6).
Query upper limit.
Path to configure lower limit:
Set lower limit (-100e6 to
100e6).
Query lower limit.
Enable or disable limit test.
Query state of limit test.
Query test result (1=pass,
0=fail).
Path to clear failed test:
Clear failed test indication.
Enable or disable auto clear.
Query auto clear.
Re-perform limit tests.
ON
2
-2
OFF
ON
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
*Note: ON is the *RST default parameter, and OFF is the :SYSTem: PRESet default.
Table 7. DISPlay command summary
Command
Description
Default
param.
SCPI
:DISPlay
[:WINDow[1]]
:TEXT
:DATA <a>
:DATA?
:STATe <b>
:STATe?
:ENABle <b>
:ENABle?
Path to control user text mes-
sages.
Define ASCII message “a”
(up to 12 characters).
Query text message.
Enable or disable
message mode.
Query text message state.
Enable or disable the front panel
display.
Query state of the display.
(Note 1)
(Note 2)
(Note 3)
√
√
√
√
√
√
√
√
Notes:
1. *RST and :SYSTem:PRESet have no effect on a user defined message. Cycling
power cancels all user defined messages.
2. *RST and :SYSTem:PRESet have no effect on the state of the message mode.
Cycling power disables (OFF) the message mode.
3. *RST and :SYSTem:PRESet have no effect on the display circuitry. Cycling power
enables (ON) the display circuitry.
Table 6. CALCulate command summary (cont.)
Command
Description
Default
param.
SCPI