15
While in 2-wire or 4-wire ohms (
Ω
2 or
Ω
4), select offset
compensation by pressing SHIFT and then O COMP.
Ratio
This feature lets you perform a ratio calculation
between the sense input (denominator) and the measure
input (numerator) for DC voltage. It lets you compare
DC voltages (measure input) to a reference voltage
(sense input). Note that SENSE LO and LO should be
connected together.
CAUTION
Do note apply more than 1000V peak to the measure
INPUT terminals or more than 350V peak to the SENSE
terminals, or instrument damage may occur.
While in DC volts (DCV), select ratio by pressing SHIFT
and then RATIO. Use the
,
,
▲
, and
▼
keys to tog-
gle RATIO to ON and SENSIN to OFF. The display will
read RA for ratio.
Measure DCV using SENSE terminals
The sense terminals can be used to measure DC voltage
on the 100mV, 1V, and 10V ranges. Note that SENSE LO
and LO should be connected together.
CAUTION
Do note apply more than 350V peak to the SENSE ter-
minals, or instrument damage may occur.
While in DC volts (DCV), press SHIFT and then RATIO.
Use
the , ,
▲
, and
▼
keys to toggle RATIO to OFF
and SENSIN to ON. The display will read RS for voltage
on SENSE terminals.
Triggers
Factory defaults place the instrument in a continuous
(immediate) measurement mode. The instrument can be