43
Command
Description
Default
param.
SCPI
:TCOefficient <n>
:TCOefficient?
:OFFSET <n>
:OFFSet?
Specify temp coefficient
(-0.09999 to 0.09999).
Query temp coefficient.
Specify voltage offset at 0°C
(-0.09999 to 0.09999).
Query voltage offset.
2e-4
5.463e-2
:FRTD
:TYPE <name>
:TYPE?
:RZERo <NRf>
:RZERo?
:ALPHa <NRf>
:ALPHa?
:BETA <NRf>
:BETA?
:DELTa <NRf>
:DELTa?
Path to configure FRTD sensor.
Select FRTD sensor type
(PT100, D100, F100,
PT3916, PT385, USER)
Query FRTD sensor type.
Specify RZERo value
(0 to 10,000).
Query RZERo value.
Specify ALPHa value
(0 to 0.01).
Query ALPHa value.
Specify BETA value
(0 to 1.00).
Query BETA value.
Specify DELTa value
(0 to 5.00).
Query DELTa value.
PT100
100.00
0.00385
0.11100
1.50700
:FREQuency
:THReshold
:VOLTage
:RANGe <n>
:RANGe?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
:PERiod
:THReshold
:VOLTage
:RANGe <n>
Path to configure frequency.
Path to select the threshold
voltage range:
Select threshold range
(0 to 1010).
Query threshold range.
Specify reference (0 to 1.5e7).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resolu-
tion (4 to 7).
Query resolution.
Path to configure period.
Path to select the threshold
voltage range:
Select threshold range
(0 to 1010).
10
0
OFF
7
10
Table 10. SENSe command summary (cont.)