Configuration method
Configuration method
Configuration method
Configuration method
4-10
[6--] Noise
[6-1] Noise table number
Parameter
Default Value
Description
0-10
0
Input required no. of points for table
[6-2] Noise table
(
[600] ~ [609]
)
Parameter
Default Value
Description
0.000-20.000
0.000
Noise position by distance
Parameter range in metric [m] units
If measurement unit is [ft], display will be converted to show ‘feet’ units.
[6-3] Amplitude threshold
(
[610] ~ [619]
)
Parameter
Default Value
Description
0-10000
0
[1/100dB]
If you create a noise table incorporating settings [6-1] ~ [6-3], you can produce a line chart such
as that shown below which enables you to eliminate lower signal strength noises below the line.
Amplitude at distance 0m and 20 m is set at 0 dB as default.
[6-4] Delete All
Parameter
Value
Description
0
Do not delete
1
Delete All
Setting “1” clears noise table