Troubleshooting
Troubleshooting
Troubleshooting
Troubleshooting
7-1
7 Troubleshooting
SYMPTOM
ACTION
No level reading
Check the power supply.
Incorrect level reading
Check the gauge calibration.
Check that gauge has not locked on an
interfering object.
Check for correct mechanical installation.
Serial Communication failure
Check the COM port setting in the LR setup
program.
Check the serial communication address.
Check
the
cable
and
HART
modem
connections and make sure that correct cable
and resistor are used.
Display Panel window is blank Check the power supply.
Check the status of the gauge in the LCD display.
Status descriptions are as follows.
[E00]
Normal Operation -
[E01]
Internal Error
Critical Error. Please contact service
[E02]
Memory Error
Critical Error. Please contact service
[E03]
Receive Error
No measured data. Please contact service
[E04]
Tank Table Error
Incorrect setting on tank table. Set up correctly.
[E05]
Noise Table Error Incorrect setting on noise table. Set up correctly.
[E99]
Searching Echo
Conducting echo search.