Guidance and Manufacturer’s Declaration – Electromagnetic Immunity to Interference
KBL Tower pureEnergy is intended for use in the electromagnetic environment specifi ed below. The studio
operator of the lamp KBL Tower pureEnergy should assure that it is used in such an environment.
Immunity tests
IEC 60601-
Test level
Compliance level
Electromagnetic environment
- Guidance
Electrostatic discharge accord-
ing to IEC 61000-4-2
+/- 6 kV
contact discharge
+/- 8
air discharge
+/- 6 kV
contact discharge
+/- 8
air discharge
Floors should be made of wood or concrete
or be covered with ceramic tiles. If fl oors are
covered with synthetic material, the relative
humidity needs to be at least 30 %.
Fast transient electrical distur-
bances/bursts according to IEC
61000-4-4
± 2 kV
for power-supply lines
± 1 kV
for input and output lines
± 2 kV
for power-supply lines
± 1 kV
for input and output lines
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Surge voltages (surges)
according to IEC 61000-4-5
± 1 kV
differential mode voltage
± 2 kV
common mode voltage
± 1 kV
differential mode voltage
± 2 kV
common mode voltage
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Voltage dips, short interruptions
and voltage variations on pow-
er-supply input lines
according to IEC 61000-4-11
< 5 % U for ½ period
(> 95 % dip)
40 % U for 5 periods
(60 % dip)
70 % U for 25 periods
(30 % dip)
< 5 % U for 5 s
(> 95 % dip)
< 5 % U for ½ period
(> 95 % dip)
40 % U for 5 periods
(60 % dip)
70 % U for 25 periods
(30 % dip)
< 5 % U for 5 s
(> 95 % dip)
Mains power quality should be that of a typi-
cal commercial or hospital environment.
If the studio operator of KBL Tower pureEn-
ergy requires continuous operation of the
device even in the event of interruptions
of the power supply, it is recommended
that KBL Tower pureEnergy is powered by
an uninterrupted power supply unit or by a
battery.
Power-frequency (50/60 Hz)
magnetic fi eld according to IEC
61000-4-8
3 A/m
30 A/m
Power-frequency magnetic fi elds should
be at levels characteristic of a typical loca-
tion in a typical commercial or hospital
environment.
Summary of Contents for Tower pureEnergy
Page 1: ...OPERATING INSTRUCTIONS TANNING DEVICE ...
Page 59: ...O p e r a t i n g I n s t r u c t i o n s KBL Tower pureEnergy 60 Remarks ...
Page 61: ...O p e r a t i n g I n s t r u c t i o n s KBL Tower pureEnergy 62 Remarks ...
Page 62: ...Ringstraße 24 26 56307 Dernbach Germany fon 49 0 26 89 94 26 0 fax 49 0 26 89 94 26 66 ...