3 Device upgrade
9.
Remove the wall bracket from the wall.
10.
Remove the column floor screws.
3.11 Re-assemble the device
NOTICE!
Refer to the supplied
Installation manual
for more instructions to these tasks.
1.
Move the column to a position where a CEPH device can be installed. Use the supplied installation
template to help you find the correct position for the device.
2.
Attach the column to the wall with CEPH wall bracket.
3.
Fix the column to the floor with a pivot screw.
4.
Install the carriage back on the column.
5.
Remove the supports installed during the device disassembly.
3.12 Install the CEPH arm
NOTICE!
Refer to the supplied
Installation manual
for more instructions to these tasks.
1.
Install the CEPH arm to the device.
2.
Rotate the device to preferred angle.
3.
Check that the column is level.
4.
Fix the column to the floor (4 screws).
3.13 Install the new sensor holder assembly
Install the new sensor holder assembly with movements, supplied with the upgrade kit.
ORTHOPANTOMOGRAPH
™
OP 3D
31