3 Device upgrade
3.10 Disassemble the device
Remove the carriage from the column to let you to re-position the column for CEPH installation.
NOTICE!
It is recommended to protect the floor with, for example, the device package cardboard
sheets. This prevents unwanted damage to the floor during the device disassembly.
1.
Make sure that the spring latch in the column is securely in place.
2.
Loosen the carriage locking bolts (3 pcs) and turn them aside.
3.
Have one person to hold the carriage from the upper shelf support.
4.
Push the spring latch in the column in to release the carriage.
ORTHOPANTOMOGRAPH
™
OP 3D
29