3 Device upgrade
3.4 Remove device covers
Remove all shown device covers.
Covers to be removed (in order):
1.
Lower shelf tray
2.
Patient handles
3.
Mirror
4.
Carriage side covers
5.
Upper shelf covers
6.
PAN/3D tubehead front cover with the
tubehead plate
7.
PAN/3D tubehead back cover
8.
Rotating unit side covers
9.
Sensor collars and sensor covers *
* Sensor collars and covers are different between 3D and PAN only devices
3.4.1 Removing the lower shelf tray
1.
Insert a screwdriver or similar into the gap between the lower shelf tray and the carriage and push
onto the dents inside cover. Do this on both sides of the lower shelf tray to loosen the tray.
2.
Turn the protruding parts of the lower shelf tray slightly and remove the lower shelf tray.
ORTHOPANTOMOGRAPH
™
OP 3D
9