3 Device upgrade
17.
Insert the compression spring to the linear movement screw.
18.
Attach and tighten the linear movement screw to the motor coupler.
19.
Fix the linear movement screw to the coupler with the retaining nut.
20.
Check that the linear movement screw is able to move a bit in horizontal direction.
21.
Move the linear movement back and forth by hand to make sure that the movement is smooth.
22.
Tighten the top screws (2 pcs) of the motor and the linear bearing housing.
23.
Tighten the side screw of the linear bearing housing gently, so that the linear bearing housing does
not move.
ORTHOPANTOMOGRAPH
™