3 Device upgrade
27.
Attach the rotating unit to the new linear movement sledge with 4 screws.
NOTICE!
If the rotating unit support has only 2 screw holes, then attach the rotating unit
with only 2 screws.
28.
Route the main cable set through the cable clamps and attach the clamps. The main cable set has
markings that indicate the correct cable position in relation to the clamps.
29.
Re-attach the main cable connectors to the i3000 and upper shelf.
ORTHOPANTOMOGRAPH
™
OP 3D
25