4 Installation steps
FOCUS
33
4.9 Setting the DIP switches
Remote Control
Connection Box Board
Description
2
3
4
Remote control and external switch in
parallel
on
on
Remote control and external switch in
series
off
off
Remote control only in use
off
on
External switch(es) only in use
on
off
Door switch disabled
on
Dip switch
Position
Description
1
ON
(closed)
Door signal from Connection Box is
connected to ground (Door is closed)
OFF (open)
Door signal from Connection Box is
connected to door switch (Causes
Error H3)
2
unused
3
ON
(closed)
Enables exposure signal for Remote
A (X62). Switch 5 must also be ON
OFF (open)
Disables exposure signal for Remote
A (X62)
4
ON
(closed)
Enables exposure signal for Remote
B (X63). Switch 6 must also be ON
OFF (open)
Disables exposure signal for Remote
B (X63)
5
ON
(closed)
Enables Remote A (X62)
OFF (open)
Disables Remote A (X62) (Read-only
mode)
6
ON
(closed)
Enables Remote B (X63)
OFF (open)
Disables Remote B (X63) (Read-only
mode)
Summary of Contents for FOCUS 215720
Page 2: ......
Page 4: ......
Page 6: ...ii...
Page 12: ...2 Unit description 6 FOCUS Exposure time control...
Page 55: ...6 Dimensions FOCUS 49 6 Dimensions Wide wall mount plate measures in mm...
Page 56: ...6 Dimensions 50 FOCUS Narrow wall mount plate measures in mm...
Page 57: ...6 Dimensions FOCUS 51 Remote control frame measures in mm...
Page 58: ...6 Dimensions 52 FOCUS Wall mount installation dimensions...
Page 59: ...6 Dimensions FOCUS 53...
Page 60: ...6 Dimensions 54 FOCUS...
Page 61: ...6 Dimensions FOCUS 55...
Page 62: ...6 Dimensions 56 FOCUS OpenMount installation dimensions...