4 Installation steps
14
FOCUS
1.
Install the KaVo unit adapter to the treatment
unit's light mounting pole according to the
instructions included with the adapter.
2.
Attach the long protective ground wire of the
adapter to the treatment unit according the
instructions with the adapter.
3.
Route the mains cable of FOCUS together with
the Ethernet cable from the Remote Control
into the bottom of the treatment unit's pole and
up through the KaVo unit mount adapter,
leaving approx. 15 cm/6 inches of the cables
sticking out through the adapter.
To make the routing upwards through the pole
easier, a string can first be routed down
through the KaVo unit mount adapter and down
the pole. Attach the string to the cables and
carefully pull up the cables using the string.
It is also possible to surface mount the remote
control (see section 4.8 Installing the remote
controller), in this option the Ethernet cable is
routed outside the treatment unit.