3 Pre-installation
8
FOCUS
Verify that the unit power is supplied through a
circuit breaker with current rating of max. 16 A
slow blow.
NOTICE!
Some countries and/or localities may
have regulations on fault current protection. Such
regulations must be observed.
Protective ground: Verify the protective ground.
Grounding resistance should be measured from the
screws in cone fixing element on the tube head
side.
Reach and coverage:
Refer to section 6 Dimensions for reach and
mechanical dimensions. The reach of the system is
measured from the tip of the cone with the arm
fully extended and the cone pointing back towards
the pivot of the wall plate. This measurement
method accommodates the normal positioning of
the Tube head to the patient.
This product is available with three different
horizontal arm lengths. This allows the system to
have three different areas of reach. Do not select a
longer arm than what is needed for the required
reach.
“Short” horizontal arm allows a reach of 176 cm
“Medium” horizontal arm allows a reach of 191
cm
“Long” horizontal arm allows a reach of 216 cm
Voltage:
Voltage range:
230 VAC line
207 - 253 VAC
115 VAC line
103 - 126 VAC
Voltage:
Fuse
:
230 VAC line
6,25 A Slow blow
115 VAC Line
6,25 A Slow blow
Summary of Contents for FOCUS 215720
Page 2: ......
Page 4: ......
Page 6: ...ii...
Page 12: ...2 Unit description 6 FOCUS Exposure time control...
Page 55: ...6 Dimensions FOCUS 49 6 Dimensions Wide wall mount plate measures in mm...
Page 56: ...6 Dimensions 50 FOCUS Narrow wall mount plate measures in mm...
Page 57: ...6 Dimensions FOCUS 51 Remote control frame measures in mm...
Page 58: ...6 Dimensions 52 FOCUS Wall mount installation dimensions...
Page 59: ...6 Dimensions FOCUS 53...
Page 60: ...6 Dimensions 54 FOCUS...
Page 61: ...6 Dimensions FOCUS 55...
Page 62: ...6 Dimensions 56 FOCUS OpenMount installation dimensions...