Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
10-8
11. As shown below, the Phantom Platform must appear below the
Field of View and the Phantom must be centered.
12. Adjust the Phantom Platform to achieve the proper height.
13. To move the phantom to the right or left, use the
Back / Front
feature on the Volume tab and then return to the Calibration tab.
If required, make adjustments and click
Preview
again. Repeat
as required.
14. Click
Capture
to start the scan process.
15. When the scan is completed, select
NO
when prompted to
automatically detect arches.
16. Select
Cancel
at the
Contourline Setup
prompt.