Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
12-2
Image Detector:
Amorphous Silicon Flat Panel (readable area), 9.37”
width x 7.56” height (23.8 cm width x 19.2 cm height)
Sensor Front Panel Attenuation Value:
Less than 1mm of aluminum
equivalent
Gray Scale:
14 bit
Voxel Size:
0.4/0.3/0.25/0.2 mm
Image Acquisition:
Single 360 degree rotation (maximum)
Scan Time:
27/9 seconds
Field of View:
(standard) 6.50 inches diameter x 5.33 inches height
(maximum) (16.5 cm diameter x 13.5 cm height (maximum)
Extended Field of View:
(optional) 9.15 inches diameter x 6.67 inches
height maximum) (23.2 cm diameter x 17 cm height (maximum))
Note:
Maximum values can be collimated down.
Primary Reconstruction:
Less than 2 minutes for 27 sec scan @ 0.4
voxel
Secondary Reconstruction:
Real Time
Power Requirements
Line Voltage:
115VAC, 200VAC or 230VAC (Factory Set)
Line Voltage Regulation requirement:
+ 10%
Line Current:
10 Amps (115V), 7.5 Amps (200V) or 5 Amps (230V)
Line Frequency:
50 Hz / 60 Hz
Phase:
Single
Main Circuit Breaker:
10 Amps (115V), 7 Amps (200V) or 5 Amps
(230V)
Power Cords
Power Cords are not supplied with the system. Three Power Cords
are required, one each for the Gantry, Computer, and Monitor.
Power Cord specifications:
10 Amp, 220-240VAC, 50 Hz / 60 Hz, 16/3 Gauge (minimum)
Length 6 to 12 ft. (2m to 4m), IEC-320-C13 Connector End.
Summary of Contents for 3D eXam
Page 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Page 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Page 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Page 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Page 161: ...k990400 September 19 2007 B 7...
Page 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Page 163: ...k990400 September 19 2007 B 9...
Page 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...