Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
10-18
Uniformity Test
18. Use the Noise Level Test dataset (previous procedure), if not
available, perform steps 1 to 13.
If using the Noise Level Test, right click the HU data (upper
right corner of image) and select
Remove all measurements
.
19. Right click the image and select
HU Statistics
from the menu.
Use the Region Tool to define a ROI (Region of Interest) in four
areas of Water, as shown above.
NOTE:
Four regions is the number of regions that can be displayed
at one time. Creating another region (fifth) will overwrite the first.