93
11 Technical data
11.1
Digital interface
11.2
Analog temperature input
11.3
Sampling interval
11.4
Measuring circuit monitoring
11.5
Binary input
11.6
Controller
Main input
Display range
Accuracy
Temperature influ-
ence
Baud rate
RS485 interface with Mod-
bus master function
0.000 to 9.999
00.00 to 99.99
000.0 to 999.9
0000 to 9999
Depends on
the sensor
used
Depends on
the sensor
used
2400 Bd
4800 Bd
9600 Bd
19200 Bd
38400 Bd
Auxiliary input
Measuring range
Accuracy
Temperature influence
Temperature Pt100
(automatic detection)
-50 to +250 °C
a
± 0,5_K (up to 100 °C)
± 0,8_K (as of 100 °C)
0.05 %/10 K
Temperature Pt1000
(automatic detection)
± 0,5_K (up to 100 °C)
± 1,0_K (as of 100 °C)
Temperature NTC/PTC
Max. 4 kOhm
Input via table with 20 pairs
of values via setup program
0.3 %
b
0.05 %/10 K
a
Can be changed to °F.
b
Depending on the supporting points.
Analog
temperature input
Digital interface with
JUMO
ecoLine
O-DO
JUMO
digiLine
O-DO S10
JUMO
tecLine
20263x
JUMO
ecoLine NTU
freely configu-
rable sensor
500 ms
Adjustable,
1 to 60 s
3 s
500 ms
1 s
Adjustable,
1 to 60 s
Input
Underrange/
overrange
Short-circuit
Open circuit
Temperature
Yes
Yes
Yes
Activation
Via potential-free contact
Function
Key lock/HOLD/alarm stop
Controller type
Limit value controller, pulse length controller, pulse frequency controller,
three-step controller, continuous controller
Controller structure
P/PI/PD/PID
Summary of Contents for AQUIS 500 RS
Page 5: ...Contents 12 5 Template for panel cut out 114 13 China RoHS 116 14 Index 117...
Page 8: ...2 Description 8 Block diagram...
Page 16: ...4 Mounting 16...
Page 112: ...12 Appendix 112 12 4 2 Manual configuration of a sensor 1 Only for temperature sensor MODBUS...
Page 113: ...113 12 Appendix...
Page 114: ...12 Appendix 114 12 5 Template for panel cut out...
Page 115: ...115 12 Appendix...
Page 119: ......