
8 Calibrating sensors
64
8.1.2
End value calibration
Requirements
• The device is supplied with voltage,
see chapter 5 "Installation", page 17 et seqq.
• The sensor is connected
• Sensor startup on the device is completed
• Calibration has been enabled, see chapter 6.7 "Administrator level", page
• The device is in "Measurement mode"
Selecting calibration methods
✱
Initiate calibration using the
key (or via the administrator level).
✱
Select "FINAL VALUE" and start with the
key.
✱
Now remove the calibration sensor from the process.
Calibration for 100% saturation (end value)
✱
Bring the sensor to the defined 100% saturation state by placing it in
water-vapor-saturated air.
CAUTION!
The response of the outputs during calibration depends on their configuration.
Summary of Contents for AQUIS 500 RS
Page 5: ...Contents 12 5 Template for panel cut out 114 13 China RoHS 116 14 Index 117...
Page 8: ...2 Description 8 Block diagram...
Page 16: ...4 Mounting 16...
Page 112: ...12 Appendix 112 12 4 2 Manual configuration of a sensor 1 Only for temperature sensor MODBUS...
Page 113: ...113 12 Appendix...
Page 114: ...12 Appendix 114 12 5 Template for panel cut out...
Page 115: ...115 12 Appendix...
Page 119: ......