71
8 Calibrating sensors
8.2.2
End value calibration
Requirements
• The device is supplied with voltage,
see chapter 5 "Installation", page 17 et seqq.
• The sensor is connected
• Sensor startup on the device is completed
• Calibration has been enabled, see chapter 6.7 "Administrator level", page
• The device is in "Measurement mode"
Selecting calibration methods
✱
Initiate calibration using the
key (or via the administrator level).
✱
Select "FINAL VALUE" and confirm using the
key.
✱
Now remove the calibration sensor from the process.
✱
Bring the sensor to the defined 100% saturation state by placing it in
water-vapor-saturated air.
CAUTION!
The response of the outputs during calibration depends on their configuration.