7 Startup
"Step-by-step" instructions for sensor startup using automatic sensor configuration
Accessing the administrator level
The device is in measurement mode (normal display).
✱
Press the
key for more than 2 seconds.
✱
Use the
key to select "ADMINISTR. LEVEL".
✱
Confirm with the
key.
Enter password
✱
Use the
key to set the flashing value on the display under "PASS-
WORD" to "300" (holding down the key increases the value continuously;
pressing the
key decreases the value).
✱
Confirm with the
key.
Accessing the basic settings
✱
Use the
key to select "BASIC SETTINGS".
✱
Confirm with the
key.
✱
Use the
or
key to respond to "AUTOMATIC CONFIG. OF THE SENSOR" with
"YES" (flashing).
✱
Confirm with the
key.
The AQUIS 500 scans the Modbus-RS485 interface. If it is connected cor-
rectly, the following display appears after a short time:
✱
Confirm with the
key.
If you do not know where the device display is in the menu structure, use the
key to navigate to the next menu level up. Pressing several times will take
you back to measurement mode (normal display).