6 Operation
6.10
MANUAL mode / simulation mode
This function can be used to set the switching outputs and the analog outputs
of the device to a defined status manually. The makes dry startup, trouble-
shooting, and servicing easier, for example.
Simulation mode accesses switching outputs K1/K2 and analog outputs A1/
A2 directly. If simulation mode has been selected, MANUAL mode cannot be
used.
In MANUAL mode, the settings of the "higher order controllers" are taken into
account.
6.10.1 MANUAL mode using "higher order control functions"
Higher order switching functions
The JUMO AQUIS 500 is configured to higher order controller functions if
the following settings are made:
OPERATOR LEVEL > CTRL. CHAN. 1 OR 2 > CONTROLLER TYPE: LIMIT or
PULSE WIDTH or PULSE FREQ. or CONTINUOUS or MODULATING
Depending on the configured controller type, manual mode affects the analog
outputs and/or switching outputs. For recommended procedure, see chapter
6.10.3 "Simulation of analog outputs using MANUAL mode", page 40.
Selecting manual mode
NOTE!
The MANUAL mode parameter is locked in the device per default, which
means that it can only be activated by an administrator.
For other operators, the parameter must be enabled first – see chapter 6.7.3
"Enable level", page 31.
Summary of Contents for AQUIS 500 RS
Page 5: ...Contents 12 5 Template for panel cut out 114 13 China RoHS 116 14 Index 117...
Page 8: ...2 Description 8 Block diagram...
Page 16: ...4 Mounting 16...
Page 112: ...12 Appendix 112 12 4 2 Manual configuration of a sensor 1 Only for temperature sensor MODBUS...
Page 113: ...113 12 Appendix...
Page 114: ...12 Appendix 114 12 5 Template for panel cut out...
Page 115: ...115 12 Appendix...
Page 119: ......