90
10 Remedying errors and malfunctions
10.1
General error messages
Problem/error message
Possible cause
Measure
No measured value dis-
play or
current output
No voltage supply
Check voltage supply
Measured value display
0000 or
current output 4 mA
Sensor not submerged
in medium
Container level too low
Top-up container
Flow fitting blocked
Clean flow fitting
Sensor measurand
unsuitable or sensor faulty
Clean sensor
Replace sensor
Incorrect or
unstable
measured value display
Sensor measurand
unsuitable
Clean sensor
Calibrate sensor
Replace sensor
Sensor not positioned correctly
Select different installation loca-
tion
No through-mixing
Ensure homogeneous medium.
Ensure sensor has
rinsing flow on all sides
Air bubbles disrupt the measure-
ment
Optimize mounting
Measured value display
8888,
Temperature display "ok",
flashing
Overrange
Select suitable measuring range
Connected sensor faulty
Replace connected
sensor
Measured value display
8888,
Temperature display 8888
flashing
Temperature overrange or under-
range
The temperature of the medium
must be within the admissible
range.
Replace sensor
Temperature probe short circuit or
open circuit
Replace sensor or cable
Line break
Replace sensor or cable
No sensor connected
Connect sensor
Configure sensor on device
Short circuit
- Cable
- Sensor
- Terminals
Check cables and connections.
Replace sensor.
Temperature too high
Keep within admissible range
Summary of Contents for AQUIS 500 RS
Page 5: ...Contents 12 5 Template for panel cut out 114 13 China RoHS 116 14 Index 117...
Page 8: ...2 Description 8 Block diagram...
Page 16: ...4 Mounting 16...
Page 112: ...12 Appendix 112 12 4 2 Manual configuration of a sensor 1 Only for temperature sensor MODBUS...
Page 113: ...113 12 Appendix...
Page 114: ...12 Appendix 114 12 5 Template for panel cut out...
Page 115: ...115 12 Appendix...
Page 119: ......