8 Calibrating sensors
✱
Wait until the displayed value is stable at "100.0% Sat." or has stabilized
near this value.
✱
Confirm with the
key.
The slope determined by the device (deviation from the factory calibration) is
displayed along with the corrected measured value, which should be 100.0%
Sat. immediately after acknowledgement.
✱
Now reinstall the sensor in the process.
✱
Use the
key to apply the calibration or the
key to reject the calibra-
tion.
The device returns to measurement mode.
NOTE
Water-vapor-saturated air occurs, for example, directly above the water sur-
face in a half-full glass of water at room temperature.
Make sure there are no drops of water on membrane. The sensor must remain
dry during measurement. The use of a holding device (e.g. a test tube) is rec-
ommended.