10 Remedying errors and malfunctions
92
10.2
JUMO digiLine O-DO S10 sensor error messages
10.2.1 Calibration
When an error occurs, the device checks after 30 seconds of waiting time whether the error condi-
tions still exist. If this is the case, a further check is performed after an additional 30 seconds. If the
error conditions still exist after a total of 5 intervals of 30 seconds each, the device aborts the cali-
bration process.
Error message
Possible cause
Measure
Measured oxygen value too low
for calibration
Check measurement configuration
Repeat calibration
Measured oxygen value too high
for calibration
Check measurement configuration
Repeat calibration
Temperature of the measurement
medium < 5 °C
Adhere to admissible temperature
range
Temperature of the measurement
medium > 40 °C
Adhere to admissible temperature
range
Measurement medium tempera-
ture fluctuation too great for cali-
bration process
Keep temperature stable during
calibration
Entered reference value too high
for the actual oxygen content
Check oxygen content
Repeat calibration
Entered reference value too low for
the actual oxygen content
Check oxygen content
Repeat calibration
Measured oxygen value fluctua-
tion too great for calibration pro-
cess
Wait until the measured value has
stabilized.
Device has aborted the calibration
process after 5 waiting loops
1
.
Check conditions for calibration,
see chapter 8 "Calibrating sen-
sors", page 63
Repeat calibration
1
When an error occurs, the device checks after 30 seconds of waiting time whether the error conditions still
exist. If this is the case, a further check is performed after an additional 30 seconds. If the error conditions
still exist after a total of 5 intervals of 30 seconds each, the device aborts the calibration process.
Summary of Contents for AQUIS 500 RS
Page 5: ...Contents 12 5 Template for panel cut out 114 13 China RoHS 116 14 Index 117...
Page 8: ...2 Description 8 Block diagram...
Page 16: ...4 Mounting 16...
Page 112: ...12 Appendix 112 12 4 2 Manual configuration of a sensor 1 Only for temperature sensor MODBUS...
Page 113: ...113 12 Appendix...
Page 114: ...12 Appendix 114 12 5 Template for panel cut out...
Page 115: ...115 12 Appendix...
Page 119: ......