7 Startup
54
Electrical connection of sensor type 202630
1
with digital interface
and flow monitor
1
Also applies for the types 202631, 202634, 202636 and 202637 with a digital interface.
(3)
(4)
Terminal assignment for flow monitor
Wire color
Function
Connection
Brown
+12 V
Row 1, terminal 11
Black
Contact
(NPN nor-
mally open
contact)
Row 2, terminal 12
Blue
GND
Row 1, terminal 12
(5)
(1)
( 2)
Terminal assignment of JUMO digiLine master con-
necting cable
Pin M12
(wire color)
Function
Connection
5 (gray)
RS485 +
Row 2, terminal 4
4 (black)
RS485 -
Row 2, terminal 3
3 (blue)
GND
Row 2, terminal 1
2 (white)
+24 V
Row 1, terminal 14
1 not
connected
(black)
Shield
Row 2, terminal 7
(1)
Terminal block 1
(2)
Terminal block 2
(3)
JUMO digiLine master connecting cable for 705001 with exposed wire ends at one end for connec-
tion to devices with screw or spring-cage terminals
(4)
JUMO tecLine Cl2 (sensor for free chlorine) with digital interface, type 202630/50 or /53...
(5)
JUMO individual fitting flow monitor, type 202811/20...