8 Calibrating sensors
72
Start end value calibration
✱
Wait until the displayed value has stabilized.
✱
Start calibration using the
key.
The "ENTRY REFERENCE" display flashes and shows the last determined
oxygen saturation value
NOTE
Water-vapor-saturated air occurs, for example, directly above the surface in a
half-full sample vessel of tap water.
To achieve successful calibration, the following points must be taken into
account:
• Both the sample vessel and the water must be clean.
• The water should be left to stand in the sample vessel for at least 30 min-
utes before calibration, so that an equilibrium is established in the gas
exchange between water and ambient air.
• Ideally, a holder should be installed to mount the sensor above the water
surface.
• The sensor must be kept dry during the calibration process. Drops of water
adhering to the sensor membrane could distort the measurement result.
• The air pressure and, above all, the temperature must remain constant
during the calibration.
• Direct sunlight and drafts in the immediate vicinity of the sample vessel
should be avoided.
• Wait a few minutes until the measured temperature stops changing. In the
event of larger temperature differences between water and air, this can
take up to 30 minutes.
Summary of Contents for AQUIS 500 RS
Page 5: ...Contents 12 5 Template for panel cut out 114 13 China RoHS 116 14 Index 117...
Page 8: ...2 Description 8 Block diagram...
Page 16: ...4 Mounting 16...
Page 112: ...12 Appendix 112 12 4 2 Manual configuration of a sensor 1 Only for temperature sensor MODBUS...
Page 113: ...113 12 Appendix...
Page 114: ...12 Appendix 114 12 5 Template for panel cut out...
Page 115: ...115 12 Appendix...
Page 119: ......