19
11 Calibration in general
Assessment criteria
pH calibrations
(glass electrodes and ISFET connected to analysis measuring inputs as well as standard sig-
nals connected to universal inputs)
pH calibrations (antimony electrodes connected to analysis measuring inputs)
Redox zero-point calibration
Ammonia calibration
Calibration value [unit]
—
—
Zero point [pH]
...
<
5
≤
...
<
6 to 8
<
...
≤
9
<
...
Slope [%]
...
<
75
≤
...
<
89.6 to 103.1
<
...
≤
110
<
...
Calibration value [unit]
—
—
Zero point [pH]
...
<
-2 to +2
<
...
Slope [%]
...
<
10 to 110
<
...
Calibration value [unit]
—
—
Zero point [mV]
...
<
-200
≤
...
<
-120 to +120
<
...
≤
+200
<
...
NOTE!
There is no assessment of the calibration values in the case of a Redox 2-point
calibration.
Calibration value [unit]
—
—
Zero point [mV]
...
<
-612
≤
...
<
-312 to +588
<
...
≤
+888
<
...