11 Calibration in general
20
Calibration of conductivity sensors
(analysis measuring inputs and standard signals connected to universal inputs)
Calibration value [unit]
—
—
Rel. cell constant (CR) [%]
...
<
50
≤
...
<
75 to 125
<
...
≤
150
<
...
Rel. cell constant (Ci) [%]
...
<
80
≤
...
<
90 to 110
<
...
≤
120
<
...
Calibration value [unit]
—
—
Temperature coefficient
(CR) [%/K]
...
<
0 to 8
<
...
Temperature coefficient
(Ci) [%/K]
...
<
0 to 5.5
<
...
NOTE!
No assessment of the calibration values is performed for universal inputs in the
"linear scaling" operation mode.