11 Calibration in general
16
•
You must be logged in as a user with the right to perform calibrations. Fac-
tory-preset users have all of these rights.
v
„Passwords and user rights“, Page 61
•
You must ensure that the calibration default settings for the individual analy-
sis inputs and, possibly, the universal inputs are set correctly. The explana-
tions of the calibration default settings can be found in the chapters on
calibration of the various analysis sensors.
•
In the case of Ci analysis inputs, it should be noted that the optional boards
must undergo a basic calibration during the initial commissioning. If this has
not yet been performed, it is necessary to do so prior to any other calibration.
v
Chapter 10.2 "Ci base calibration", Page 10
Manual entry of calibration values
If calibration values are known, they can also be entered manually. This may be
the case with temperature-compensated conductivity measurements, for in-
stance, when the temperature coefficient of a liquid is known. Known calibration
values are entered manually under:
Device Menu
r
Calibration
r
Select Analog Input
r
Calibration Values
NOTE!
Incorrectly entered calibration values result in incorrect measuring values.
Correct measurements are essential for control systems and limit
monitoring.
Summary of Contents for 202580
Page 6: ...Content...
Page 28: ...12 Calibrating a pH measuring chain 28...
Page 42: ...15 Calibrating CR conductivity sensors 42...
Page 52: ...16 Calibrating Ci conductivity sensors 52...
Page 60: ...17 Calibrating universal inputs 60...
Page 79: ...79 18 Technical data...
Page 80: ...18 Technical data 80...
Page 81: ......