45
16 Calibrating Ci conductivity sensors
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
Calibration default settings for calibrating the relative cell constant
Calibration default settings for calibrating the temperature coefficient
Sample screen:
Ci calibration default
settings
Parameter
Setting options
Explanation
Reference conductivi-
ty
0 to 9999 mS/cm
Conductivity of the reference solution
Parameter
Setting options
Explanation
Temperature com-
pensation
Selection from
analog selection
Temperature input for automatic sensing of the test/
sample solution temperature during the calibration
Reference tempera-
ture
-50 to 150 °C
The conductivities of a sample solution at the refer-
ence temperature and operation temperature are
sensed during the calibration procedure. This yields
2 value pairs (temperature/conductivity). These val-
ue pairs provide the basis for calculating the tem-
perature coefficient.
The operation temp. must differ from the reference
temperature by at least 5 °C.
Operating tempera-
ture
-50 to 150 °C
Summary of Contents for 202580
Page 6: ...Content...
Page 28: ...12 Calibrating a pH measuring chain 28...
Page 42: ...15 Calibrating CR conductivity sensors 42...
Page 52: ...16 Calibrating Ci conductivity sensors 52...
Page 60: ...17 Calibrating universal inputs 60...
Page 79: ...79 18 Technical data...
Page 80: ...18 Technical data 80...
Page 81: ......