15 Calibrating CR conductivity sensors
4
•
with temperature sensing
A prerequisite is that temperature compensation was specified in
the calibration default settings.
Bring the temperature of the sample solution to the requested val-
ues of the reference and operation temperatures in succession.
The order does not matter. Acquisition of the individual values
takes place automatically.
•
without temperature sensing
If temperature compensation was not specified in the calibration
default settings, you must control acquisition of the value manually.
Bring the temperature of the sample solution first to the value of the
reference temperature and confirm by pressing "OK".
Then proceed in the same manner for the operation temperature.
5
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6
Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
Step
Action
requested
Temperature values
current actual value of
temperature
Display after
acquisition of first
value
remaining
requested
temperature value