37
15 Calibrating CR conductivity sensors
Calibration default settings for calibrating the relative cell constant
Calibration default settings for calibrating the temperature coefficient
Parameter
Setting options
Explanation
Reference conductivi-
ty
0 to 9999 mS/cm
Conductivity of the reference solution
Parameter
Setting options
Explanation
Temperature com-
pensation
Selection from
analog selection
Temperature input for automatic sensing of the test/
sample solution temperature during the calibration
Reference tempera-
ture
-50 to +150° C
The conductivities of a sample solution at the refer-
ence temperature and operation temperature are
sensed during the calibration procedure. This yields
2 value pairs (temperature/conductivity). These val-
ue pairs provide the basis for calculating the tem-
perature coefficient.
The operation temp. must differ from the reference
temperature by at least 5 °C.
Operation tempera-
ture
-50 to +150° C