17 Calibrating universal inputs
56
•
free chlorine, ph/Temp.-compensated
In the universal inputs calibration default settings for the "free chlorine, ph/T-
compensated" operation mode, the slope calibration is enabled and pre-config-
ured as the only available calibration routine.
Additional calibration default settings for calibration of chlorine sensors are ex-
plained in the following table.
Sample screen:
Universal input calibration
default settings in the
"free chlorine, pH/T-com-
pensated" operation
mode
Parameter
Possible settings
Explanation
Temperature com-
pensation
Selection from Analog selection
Temperature input for automatic sens-
ing of the test/sample solution tempera-
ture during the calibration
pH compensation
source
Selection from Analog selection
pH measuring input for automatic sens-
ing of the test/sample solution pH value
during the calibration
NOTE!
The "Calibration Default Settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
v
Summary of Contents for 202580
Page 6: ...Content...
Page 28: ...12 Calibrating a pH measuring chain 28...
Page 42: ...15 Calibrating CR conductivity sensors 42...
Page 52: ...16 Calibrating Ci conductivity sensors 52...
Page 60: ...17 Calibrating universal inputs 60...
Page 79: ...79 18 Technical data...
Page 80: ...18 Technical data 80...
Page 81: ......