Chapter 13 - Generator Calibration
Calibration Two (Read Current Calibration)
This section describes the requirements for calibrating the monitored (read) current level of the GEN04
Generator.
Note: “Power up into BME Mode”
includes taking the following steps – which are also contained in the service
manual (included here for reference):
•
Power down unit
•
Wait 5 seconds
•
Power up Unit while simultaneously holding the “STANDBY” and “DOWN ARROW” keys
Procedure:
1
Power up the generator into BME Mode
2
Scroll to page 2.
3
Record the ‘Current Setpoint’ value.
_______________mA
4
By specification, 75% of the value recorded in step 3 is the anticipated Level 3 current. Calculate and record
this value by
multiplying
the step 3 value above, by
0.75.
_______________mA
5
Put the generator into BME Ready State by pressing the ‘STANDBY’ Key.
6
Ensure the test tip is prepared to activate in air and is not in contact with anything during activation.
7
Activate at level 3 by pressing the “MIN” footswitch or “MIN” hand switching adapter buttons.
8
Continue activating the generator at level 3 for at least 1 second after the pre-run test is complete.
9
Deactivate the hand piece by releasing footswitch or hand switching adapter buttons..
10
Put the generator into BME Standby State by pressing the ‘STANDBY’ Key.
11
Scroll to page 18.
12
Press the ‘TEST’ key one time (only one time) to accept the calibration current value.
13
The text “Value Accepted!” should flash if the calibration procedure was successful.
Fail_____ Pass_____
14
Record the current displayed on the screen.
_______________mA
15
Power the unit off for at least 5 seconds.
16
Power the unit into BME Mode.
17
Enter BME Ready State by pressing the “STANDBY” Key.
18
Ensure the test tip is prepared to activate in air and is not in contact with anything during activation.
19
Activate at level 3 by pressing the “MIN” footswitch or “MIN” hand switching adapter buttons.
20
Continue activating the generator at level 3 for at least 1 second after the pre-run test is complete.
21
Deactivate the hand piece by releasing footswitch or hand switching adapter buttons.
22
Go to BME Standby State by Pressing the “STANDBY” Key.
23
Scroll to page 18.
24
Record the ‘I (LEVEL 3)’ current displayed on page 18.
_______________mA
25
Verify that the reading is within 1% of the value calculated in step 4. If not, carefully re-execute both
CALIBRATION ONE & CALIBRATION TWO procedures again (up to three times.)
Fail_____ Pass_____
26
Power OFF the Generator.
GEN04 Serial Number:______________________________
Technician Signature:______________________________________________Date:__________________________
Summary of Contents for ETHICON ENDO-SURGERY HARMONIC 300
Page 1: ...HARMONIC Generator 300 System Service Manual ...
Page 2: ...GEN04 ...
Page 6: ...4 GEN04 Service Manual ...
Page 10: ...8 GEN04 Service Manual ...
Page 16: ...14 GEN04 Service Manual ...
Page 24: ...22 GEN04 Service Manual ...
Page 26: ...24 GEN04 Service Manual ...
Page 28: ...26 GEN04 Service Manual ...
Page 44: ...42 GEN04 Service Manual ...
Page 50: ...48 GEN04 Service Manual Power Supply Fig 12 5 Power Supply Troubleshooting Flowchart ...
Page 52: ...50 GEN04 Service Manual Front Bezel Fig 12 7 Front Bezel Troubleshooting Flowchart ...
Page 58: ...56 GEN04 Service Manual ...
Page 62: ...60 GEN04 Service Manual ...
Page 66: ...64 GEN04 Service Manual ...
Page 68: ...66 GEN04 Service Manual ...
Page 72: ...70 GEN04 Service Manual ...
Page 76: ...74 GEN04 Service Manual ...
Page 77: ...75 GEN04 Appendix B GEN04 Main Assembly Drawing ...
Page 79: ...77 GEN04 Appendix C GEN04 Top Level Diagram Hand Piece ID Hand Piece Transducer ...