38
GEN04
Service Manual
ii
.
Record the value of |Zr|
- |Zr| is the value of the hand piece impedance as
measured by the generator, and for a good hand piece and test tip, this value should
be between 15 and 46 Ohms.
Note: As the hand piece ages, the |Zr| value increases; therefore, if the |Zr| value is
above 40 Ohms the hand piece is approaching its “end of useful life.” Using a hand piece that
operates near its end of useful life may result in one or more of the following:
•
generator error codes
•
hand piece over heating
•
generator lock outs
|Zr| Value = ________________________________________________
8
Press the standby button to place generator in “
Ready State.
”
9
Activate at level 5 until the pre-run test is complete.
10
Stop activation after pre-run test completes
11
Reactivate at level 5.
12
Record
the value of “
impedance
” on the display while continuing to activating at level 5.
Impedance = _______________________________________________
Note: As the hand piece ages, the impedance value increases; therefore, if the impedance
value is above 40 Ohms the hand piece is approaching its “end of useful life.” Using a hand
piece that operates near its end of useful life may result in one or more of the following:
•
generator error codes
•
hand piece over heating
•
generator lock outs
If you need further assistance, please contact the Ethicon Endo-Surgery Customer Support Center at 1-800-USE-ENDO
(800-873-3636), Option 2.
Summary of Contents for ETHICON ENDO-SURGERY HARMONIC 300
Page 1: ...HARMONIC Generator 300 System Service Manual ...
Page 2: ...GEN04 ...
Page 6: ...4 GEN04 Service Manual ...
Page 10: ...8 GEN04 Service Manual ...
Page 16: ...14 GEN04 Service Manual ...
Page 24: ...22 GEN04 Service Manual ...
Page 26: ...24 GEN04 Service Manual ...
Page 28: ...26 GEN04 Service Manual ...
Page 44: ...42 GEN04 Service Manual ...
Page 50: ...48 GEN04 Service Manual Power Supply Fig 12 5 Power Supply Troubleshooting Flowchart ...
Page 52: ...50 GEN04 Service Manual Front Bezel Fig 12 7 Front Bezel Troubleshooting Flowchart ...
Page 58: ...56 GEN04 Service Manual ...
Page 62: ...60 GEN04 Service Manual ...
Page 66: ...64 GEN04 Service Manual ...
Page 68: ...66 GEN04 Service Manual ...
Page 72: ...70 GEN04 Service Manual ...
Page 76: ...74 GEN04 Service Manual ...
Page 77: ...75 GEN04 Appendix B GEN04 Main Assembly Drawing ...
Page 79: ...77 GEN04 Appendix C GEN04 Top Level Diagram Hand Piece ID Hand Piece Transducer ...