36
GEN04
Service Manual
Error Code 7: Hand Switch
Error Code 7 indicates the hand switch is stuck in the ON position.
Confirm contacts in the distal end of hand piece and in the proximal
end of the hand switching adaptor are dry or replace the hand
switching adaptor.
Note: If the error persists, replace hand switch.
Fig. 9-7 Hand Activation Assembly Error Screens