43
GEN04
Chapter 12 - Hardware Troubleshooting
The procedures described on the following pages require the generator to be powered up while the cover is removed.
Caution should be taken to avoid making direct contact with any of the internal components while the unit is powered up.
Warning:
The AC mains voltage presents a shock hazard and is accessible when the generator cover is removed.
Caution must be used near internal system components, which attach directly to the AC mains such as the power entry
module, mains fuses, power switch, and open-frame 48V power supply.
Test Equipment and Tools
High-level troubleshooting of the Harmonic Scalpel Generator requires the following tools:
•
Digital Multi-meter
•
Phillips screwdriver
Fig. 12-1 Front Bezel Components and Cooling Fan
Cooling Fan
Hand Piece
Receptacle
Front Bezel
LCD
Display
Summary of Contents for ETHICON ENDO-SURGERY HARMONIC 300
Page 1: ...HARMONIC Generator 300 System Service Manual ...
Page 2: ...GEN04 ...
Page 6: ...4 GEN04 Service Manual ...
Page 10: ...8 GEN04 Service Manual ...
Page 16: ...14 GEN04 Service Manual ...
Page 24: ...22 GEN04 Service Manual ...
Page 26: ...24 GEN04 Service Manual ...
Page 28: ...26 GEN04 Service Manual ...
Page 44: ...42 GEN04 Service Manual ...
Page 50: ...48 GEN04 Service Manual Power Supply Fig 12 5 Power Supply Troubleshooting Flowchart ...
Page 52: ...50 GEN04 Service Manual Front Bezel Fig 12 7 Front Bezel Troubleshooting Flowchart ...
Page 58: ...56 GEN04 Service Manual ...
Page 62: ...60 GEN04 Service Manual ...
Page 66: ...64 GEN04 Service Manual ...
Page 68: ...66 GEN04 Service Manual ...
Page 72: ...70 GEN04 Service Manual ...
Page 76: ...74 GEN04 Service Manual ...
Page 77: ...75 GEN04 Appendix B GEN04 Main Assembly Drawing ...
Page 79: ...77 GEN04 Appendix C GEN04 Top Level Diagram Hand Piece ID Hand Piece Transducer ...