Safety instructions
Before beginning to work with the system, all users are required to read these safety instructions.
Power supply
Power is supplied to the system via an internal medical-grade power supply.
Battery power
l
Charging – the scanner battery will be fully charged after being plugged into a power source for
2.5 hours.
l
With a fully-charged battery, you can scan up to 30 minutes.
l
The scanner is provided with two Li-ion rechargeable battery packs. There is a
danger of battery explosion if the screen is damaged. Do not use the scanner if dropped or any
damage is noticed. Contact Customer Support. Use only the original AC/DC adaptor to
charge the batteries.
l
The screen should be stored and operated following the environmental conditions that appear
in this manual. Do not expose the scanner to extreme heat sources, such as radiators and
fireplaces.
l
Never operate the device without batteries! Do not use the batteries for any purpose other
than the intended use of the product. Discard used batteries according to the manufacturer's
instructions and local requirements.
Electric
warnings
l
Do not remove external panels, covers, and batteries in order to avoid electrical shock. There
are no user-serviceable parts inside. You can open the cover of the diagnostics panel in case
of system failure, only when required by Customer Support.
l
Do not connect the scanner to a mains supply without protective grounding, in order to avoid
the risk of electrical shock.
Electric
precautions
l
Do not connect a non-Align-approved web camera to the USB sockets on the rear of the touch
screen, in order to avoid the risk of electrical shock.
l
Do not connect a power cable that is not approved by Align Technology to the system, in order
to avoid electrical shock.
Wireless LAN
The system comes equipped with a wireless LAN unit.
Safety
classifications
l
Type of protection against electrical shock: Class 1.
l
Degree of protection against electrical shock: Type BF.
l
Degree of protection against harmful ingress of water: Ordinary.
l
Equipment not suitable for use in the presence of flammable anesthetic mixtures.
l
Mode of operation: Continuous.
Prescription
health device
The system serves as a prescription medical device and should be operated by qualified health-
care providers only.
Scanner
precautions
l
The wand emits red laser light (680nm Class 1), as well as white LED emissions and 850nm
LED emissions. Normal usage of the wand does not present any danger to the human eye.
Avoid shining the wand directly into the patient’s eyes.
iTero Element® 5D Plus User manual
© 2020 Align Technology, Inc. All rights reserved.
v
Summary of Contents for Element 5D Plus
Page 1: ...iTero Element 5D Plus User manual it starts with...
Page 2: ......
Page 26: ...iTero Element 5D Plus User manual 8 2020 Align Technology Inc All rights reserved...
Page 44: ...iTero Element 5D Plus User manual 26 2020 Align Technology Inc All rights reserved...
Page 122: ...iTero Element 5D Plus User manual 104 2020 Align Technology Inc All rights reserved...
Page 130: ...iTero Element 5D Plus User manual 112 2020 Align Technology Inc All rights reserved...
Page 136: ...iTero Element 5D Plus User manual 118 2020 Align Technology Inc All rights reserved...
Page 138: ...iTero Element 5D Plus User manual 120 2020 Align Technology Inc All rights reserved...
Page 184: ...iTero Element 5D Plus User manual 166 2020 Align Technology Inc All rights reserved...
Page 188: ...iTero Element 5D Plus User manual 170 2020 Align Technology Inc All rights reserved...
Page 198: ...iTero Element 5D Plus User manual 180 2020 Align Technology Inc All rights reserved...
Page 202: ...iTero Element 5D Plus User manual 184 2020 Align Technology Inc All rights reserved...
Page 207: ......