Instructions for Replacing the Housing:
1. To re-install the user interface panel, plug the LCD cable back into the motherboard
(Figure 4).
Note orientation of red wire.
Re-attach the cannula tubing to the cannula
connector.
2. Reconnect the UIP cable to the motherboard (Figure 3).
3. Test the function of each button and the LCD. If anything does not work, reseat the
cables in the connectors and check for debris in the connectors.
4. Re-install the four screws that connect the user interface panel to the chassis (Figures 1
and 2). Tighten screws to 5 in-lbs.
5. Place the device on its back side.
6. Seat the feet of the front housing into the corresponding grooves on the chassis (Figure
11) then push down and toward the device to secure the housing in the chassis
Figure 11: Housing feet and corresponding chassis grooves
7. Place the device on its front side and repeat step 6 to secure the back housing into the
chassis.
8. Replace the two screws that connect the back housing to the chassis and the two screws
that connect the front housing to the chassis. Tighten screws to 3 in-lbs.
9. Return the concentrator to the upright position.
10. Slide the battery back into the chassis. Ensure the battery clip is fully seated.