96-06434-00-01 revA
Service Manual, Inogen One G3HF
-23-
3. Gently unplug the cannula tube and the LCD cable from the user interface panel (Figure
4). Remove the user interface panel.
Figure 4: Remove cannula tube and LCD cable from user interface panel
4. Remove the two screws from the receptacle that needs replacing (Figure 5).
Figure 5: Column Receptacle Screws
5. Remove the receptacle from the molded column tube using a small flathead screw driver, if
necessary.
6. If replacing the Short Molded Column Tube, carefully disconnect the feed waste
manifold harnesses and the product manifold harness from the motherboard using pliers
to pull the white connector off of the pins and away from the motherboard (Figures 6
and 7). Then remove the tube from the Feed Waste Manifold using a small flathead
screw driver (Figure 9).
Cannula Tube
LCD Cable