![Inogen One G3HF Service Manual Download Page 130](http://html1.mh-extra.com/html/inogen/one-g3hf/one-g3hf_service-manual_2064301130.webp)
Automobile receptacle fuse
blows or overheats
The power outlet on the automobile is not
rated to handle 120W
Check car manual for electrical
specification of the power outlet or use a
different outlet in the car
If automobile power outlet is not rated
to handle 120W then the concentrator
cannot be used with automobile power
(in this car) without a dedicated outlet
and wiring.
Auto Power
Concentrator runs and
charges battery on DC
automobile power
Automobile does not power
the concentrator
The input voltage is too low
Try a different outlet in the vehicle. If
problem persists, replace the auto power
cable with a DC power supply (BA-302)
to increase the voltage output.
Auto Power
Concentrator runs and
charges battery on DC
automobile power
Automobile does power the
concentrator, but overheating of
the cable is caused.
The power connection is shorted, causing the
DC cable to become overheated.
Send the DC automobile power cable back
to Inogen.
1
Power Fault
Expected Behavior
Abnormal Behavior
Fault
Troubleshooting & Repair
Battery Power
Concentrator displays the
Inogen logo on the LCD
screen and beeps when a
battery is inserted
Empty battery voltage =
12.0V
Full battery voltage =
16.8V
No reaction when battery is
installed
Battery stays at 0% while
external power is present
Battery Icon is flashing on the
LCD screen
The battery is depleted or not recognized by
the concentrator
Remove and reinsert the battery or
apply external power to bring the
concentrator into standby mode
Charge the battery by attaching external
power.
The battery might need 1-2 hours before
normal charging begins if the battery is
severely depleted (ie, left on a
concentrator during storage).
If battery charging never resumes, then
the battery should be replaced
Check the battery voltage by measuring
the voltage across the outside two pins
on the battery
Verify that the connector receptacle and
connector pins are not bent or damaged.
Concentrator displays
charging status and
remaining run time status
Battery time remaining says
0:00
The concentrator is not communicating
with the battery
Or the battery might be severely depleted
and in need of charging
Shut down the concentrator and then
restart it. If the problem persists, try a
different battery or a different
motherboard
96-06434-00 Service Manual, Inogen One G3HF
- 158 -