O2 Delivery Error
No Error Message
O2 Delivery Error
The system detected a breath but did not
detect a bolus delivery
Check for kinked tubing to the breath
detect sensor. Replace the breath detect
sensor or the Product manifold to
resolve the no bolus condition.
3
Abnormality
Normal Behavior
Abnormal Behavior
Fault
Troubleshooting & Repair
Noise
Concentrator should be
free of rattles and
whistles
Concentrator makes loud
noise when running
Compressor tube not seated correctly
Push the tubing onto the compressor
output so that it is all the way flush
against the end of the barb. Rotate the
tubing so that the compressor does not
hit the rear wall of the chassis
Check that the compressor mounts are
not torn
Concentrator makes a
whistling sound during part
of the operating cycle
Product manifold check valve may have
debris in it or may be misassembled
Replace the product manifold
4
Software Information
Software Information
Motherboard Motor
Driver Software
Software v1.0 is only for
older compressors that
have an 8 wire harness
(hall sensor control).
Software v1.1 and 1.2
can be used with the
older 8 wire harnesses
and newer 3 wire
harnesses.
96-06434-00 Service Manual, Inogen One G3HF