96-06434-00-01 revA
Service Manual, Inogen One G3HF
-18-
3. Gently unplug the cannula tube and the LCD cable from the user interface panel (Figure
4). Remove the user interface panel.
Figure 4: Remove cannula tube and LCD cable from user interface panel
4. Remove the two screws from the receptacle that needs replacing (Figure 5).
Figure 5: Column Receptacle Screws
5. Remove the receptacle from the molded column tube using a small flathead screw driver, if
necessary.
Instructions for Replacing the Column Receptacles:
1. Install the column receptacle into the molded column tube, ensuring that the clocking lines
on the tube are directly on top of the receptacle barb (Figure 6). Place the receptacle on top
of the column.
Cannula Tube
LCD Cable